Agilent 1290 Infinity II Flexible Pump User Manual
93
Maintenance
7
Replace the Outlet Valve
Replace the Outlet Valve
When
If Outlet valve is defective.
Tools required
p/n
Description
Wrench, 14 mm
G4220-20012
Torque wrench 2 – 25 Nm
G4220-20041
Bit Torx 10x25 mm
Parts required
p/n
Description
G4220-60028
Outlet valve
(primary pump head)
G4220-20020
Internal gold seal for Outlet Valve
Preparations
•
Switch off pump at the main power switch
•
Open the doors
•
Use an optional solvent shutoff valve or lift up solvent filters inside solvent bottles for avoiding
leakages
1
Remove the cap from the outlet valve.
2
Open the 2.5 mm hex screw at the top of the primary
pump head, which fixes the connection capillary of the
heat exchanger (1) and remove it (2). Then lift up the
capillary and remove it from the primary pump head (3).
&
'
(
Содержание G7104A
Страница 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Страница 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Страница 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Страница 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Страница 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Страница 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...
Страница 218: ...218 Agilent 1290 Infinity II Flexible Pump User Manual 11 LAN Configuration PC and User Interface Software Setup Setup...