138
Agilent 1290 Infinity II Flexible Pump User Manual
7
Maintenance
Replace Parts of the Inline Filter
Replace Parts of the Inline Filter
Tools required
p/n
Description
8710-0510
Wrench open 1/4 — 5/16 inch
Parts required
p/n
Description
5067-5407
In-Line Filter Assembly for 1290 Infinity Quaternary Pump
5067-4748
Capillary ST, 0.17 mm x 90 mm
Multi Purpose Valve to inline filter
5023-0271
Frit 0.3 µm for inline filter, 5/pk
C A U T I O N
Stuck Capillary in Multi Purpose Valve
Shortcutting the inline filter by directly connecting its right capillary to valve port 5 can
damage the Multi Purpose Valve.
The size/position of this capillary in its fitting is incompatible, so it may get stuck
irreversibly to the valve.
➔
Do not shortcut the filter by directly connecting its right capillary to valve port 5 in
case the inline filter cannot or shall not be used.
N O T E
The inline filter can be cleaned using the back-flush function in the user interface of your
Agilent instrument control software.
1
Remove the capillaries from the Multi Purpose Valve to
the inline filter.
2
Remove the inline filter from the clamp attached to the
Multi Purpose Valve.
8aVbe
>ca^cZ[^aiZg
Содержание G7104A
Страница 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Страница 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Страница 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Страница 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Страница 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Страница 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...
Страница 218: ...218 Agilent 1290 Infinity II Flexible Pump User Manual 11 LAN Configuration PC and User Interface Software Setup Setup...