120
Agilent 1290 Infinity II Flexible Pump User Manual
7
Maintenance
Pump Head Procedures
Replace Wash Seal and Gasket
1
To remove the wash seal use the steel side of the insert
tool.
2
Use the soft plastic side of the insert tool for inserting the
wash seal into the support ring.
N O T E
Clean seals before Reinstallation or use new ones.
H I N T
Store seals in isopropanol bath before installation.
Wash seal
>chZgiidda
LVh]hZVa
Hjeedgig^c\
Heg^c\
HZVa