Agilent 1290 Infinity II Flexible Pump User Manual
77
Error Information
6
Pump Error Messages
Degasser's pressure limit violation
Error ID: 29220
Pressure too far above the limit.
Seal wash pump was missing when tried to turn on
Error ID: 29223
The seal wash pump has not been detected (while being configured or
detected before)
Probable cause
Suggested actions
1
Leak in degasser chamber or degasser
tubing.
Please contact your Agilent service
representative.
2
Defect vacuum pump.
Please contact your Agilent service
representative.
3
Degasser chamber empty or connected to
air.
Block unused degasser channels.
Probable cause
Suggested actions
1
Defect cable connection to seal wash pump.
Check cable connection.
2
Defect seal wash pump motor.
Please contact your Agilent service
representative.
3
Defective main board.
Please contact your Agilent service
representative.
Содержание G7104A
Страница 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Страница 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Страница 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Страница 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Страница 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Страница 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...
Страница 218: ...218 Agilent 1290 Infinity II Flexible Pump User Manual 11 LAN Configuration PC and User Interface Software Setup Setup...