Agilent 1290 Infinity II Flexible Pump User Manual
195
LAN Configuration
11
Initialization Mode Selection
Using Stored
When initialization mode
Using Stored
is selected, the parameters are taken
from the non- volatile memory of the module. The TCP/IP connection will
be established using these parameters. The parameters were configured
previously by one of the described methods.
Figure 28
Using Stored (Principle)
Using Default
When
Using Default
is selected, the factory default parameters are taken
instead. These parameters enable a TCP/IP connection to the LAN
interface without further configuration, see
Figure 29
Using Default (Principle)
Cdc"KdaVi^aZ
G6B
HidgZY
EVgVbZiZg
6Xi^kZ
EVgVbZiZg
6Xi^kZ
EVgVbZiZg
9Z[Vjai
EVgVbZiZg
N O T E
Using the default address in your local area network may result in network problems. Take
care and change it to a valid address immediately.
Table 12
Using Default Parameters
IP address:
192.168.254.11
Subnet Mask:
255.255.255.0
Default Gateway
not specified
Содержание G7104A
Страница 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Страница 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Страница 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Страница 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Страница 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Страница 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...
Страница 218: ...218 Agilent 1290 Infinity II Flexible Pump User Manual 11 LAN Configuration PC and User Interface Software Setup Setup...