![Agilent Technologies G7104A Скачать руководство пользователя страница 174](http://html1.mh-extra.com/html/agilent-technologies/g7104a/g7104a_user-manual_2867748174.webp)
174
Agilent 1290 Infinity II Flexible Pump User Manual
10
Hardware Information
Firmware Description
Firmware Description
The firmware of the instrument consists of two independent sections:
•
a non- instrument specific section, called
resident system
•
an instrument specific section, called
main system
Resident System
This resident section of the firmware is identical for all Agilent
1100/1200/1220/1260/1290 series modules. Its properties are:
•
the complete communication capabilities (CAN, LAN, USB and RS-
232C)
•
memory management
•
ability to update the firmware of the 'main system'
Main System
Its properties are:
•
the complete communication capabilities (CAN, LAN, USB and RS-
232C)
•
memory management
•
ability to update the firmware of the 'resident system'
In addition the main system comprises the instrument functions that are
divided into common functions like
•
run synchronization through APG remote,
•
error handling,
•
diagnostic functions,
•
or module specific functions like
•
internal events such as lamp control, filter movements,
•
raw data collection and conversion to absorbance.
Содержание G7104A
Страница 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Страница 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Страница 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Страница 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Страница 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Страница 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...
Страница 218: ...218 Agilent 1290 Infinity II Flexible Pump User Manual 11 LAN Configuration PC and User Interface Software Setup Setup...