Agilent 1290 Infinity II Flexible Pump User Manual
37
Using the Pump
3
Best Practices
How to deal with solvents
•
Use clean bottles only.
•
Exchange water- based solvents daily.
•
Select solvent volume to be used up within 1 – 2 days.
•
Use only HPLC- grade solvents and water filtered through 0.2 µm filters.
•
Label bottles correctly with bottle content, and filling date / expiry
date.
•
Use solvent inlet filters.
•
Reduce risk of algae growth: use brown bottles for aqueous solvents,
avoid direct sunlight.
Select channels for Multi-Channel Gradient Valve (MCGV)
•
Use lower channels (A and/or D) for buffer solutions.
•
Regularly flush all MCGV channels with water to remove possible salt
deposits.
•
Check compatibility of buffers and organic solvents to avoid
precipitation.
Содержание G7104A
Страница 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Страница 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Страница 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Страница 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Страница 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Страница 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...
Страница 218: ...218 Agilent 1290 Infinity II Flexible Pump User Manual 11 LAN Configuration PC and User Interface Software Setup Setup...