Agilent 1290 Infinity II Flexible Pump User Manual
63
Error Information
6
Pump Error Messages
Quaternary pump shutdown during analysis
Error ID: 29199
The quaternary pump has been shut down by the control software or
control module during an analysis.
•
Parameter: 0 for off, 1 for standby.
Reading the pump encoder tag failed
Error ID: 29201
Reading the pump encoder tag has failed.
•
Parameter: 1 – 2 referring to pump drive
Probable cause
Suggested actions
1
Pump has been shut down.
Restart pump.
Probable cause
Suggested actions
1
Defect connection between encoder and
main board.
Please contact your Agilent service
representative.
2
Missing or defect tag Defect connection
between tag and encoder.
Please contact your Agilent service
representative.
Содержание G7104A
Страница 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Страница 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Страница 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Страница 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Страница 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Страница 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...
Страница 218: ...218 Agilent 1290 Infinity II Flexible Pump User Manual 11 LAN Configuration PC and User Interface Software Setup Setup...