86
Agilent E5505A User’s Guide
4
Expanding Your Measurement Experience
If the output amplitude of your DUT is not sufficient to provide an adequate
measurement noise floor, it is necessary to insert a low-noise amplifier
between the DUT and the test set. Refer to the section
page 122 for details on determining the effect the amplifier’s noise will have on
the measured noise floor.
Figure 53
Noise floor example
e5505a_user_noise_floor_ex
24 Jun 04 rev 3