300
Agilent E5505A User’s Guide
11
AM Noise Measurement Fundamentals
page 300. The spur should be adjusted such that it is between –30 and –60
dBc, with a carrier offset of 10 Hz to 20 MHz.
4
Reconnect the AM detector and measure the detector sensitivity. See
5
Turn off the spur source output.
6
Measure noise data and interpret the results.
Figure 235
Measuring relative spur level
Figure 236
Measuring detector sensitivity
DUT
E5505a_meas_relative_spur
02 Mar 04 rev 1
Calibration
source
-20 dB coupler
RF spectrum
analyzer
-10 dB
atten
DUT
E5505a_meas_detector_sen
02 Mar 04 rev 1
Calibration
source
AM detector
Test set
Noise
input
-10 dB
attenuator
-20 dB
coupler
NO T E
The quadrature meter should be at zero volts due to the blocking capacitor at the AM
detector’s output.
Содержание E1420B
Страница 24: ...24 Agilent E5505A User s Guide ...
Страница 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Страница 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Страница 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Страница 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Страница 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Страница 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Страница 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Страница 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Страница 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Страница 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Страница 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...