![Agilent Technologies E1420B Скачать руководство пользователя страница 291](http://html.mh-extra.com/html/agilent-technologies/e1420b/e1420b_user-manual_2868017291.webp)
AM Noise Measurement Fundamentals
11
Agilent E5505A User’s Guide
291
3
Locate the drive level on the AM sensitivity graph (
), and enter
the data.
4
Measure the noise data and interpret the results. The measured data will be
plotted as single-sideband AM noise in dBc/Hz.
Figure 223
AM noise calibration setup
DUT
Power meter
or spectrum
analyzer
E5505a_am_noise_cal_setup
01 Mar 04 rev 1
NO T E
The quadrature meter should be at zero volts due to the blocking capacitor at the AM
detector’s output.
Figure 224
AM detector sensitivity graph
e5505a_user_AM_cal.ai
rev2 10/24/03
Содержание E1420B
Страница 24: ...24 Agilent E5505A User s Guide ...
Страница 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Страница 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Страница 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Страница 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Страница 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Страница 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Страница 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Страница 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Страница 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Страница 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Страница 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...