Absolute Measurement Examples
6
Agilent E5505A User’s Guide
163
Selecting Loop Suppression Verification
1
Using
as a guide, navigate to the
Cal
tab.
2
In the
Cal
dialog box, check
Verify calculated phase locked loop
suppression
and
Always Show Suppression Graph
. Select
If limit is
exceeded: Show Loop Suppression Graph.
3
When you have completed these operations, click the
Close
button
.
Setup considerations for the RF synthesizer using DCFM measurement
Measurement noise floor
The signal amplitude at the test set’s R input (Signal Input) port sets the
measurement noise floor level. Use
and
determine the amplitude required to provide a noise floor level that is below
the expected noise floor of your DUT.
Figure 112
Selecting loop suppression verification
e5505a_user_select_loop
24 Jun 04 rev 3
Содержание E1420B
Страница 24: ...24 Agilent E5505A User s Guide ...
Страница 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Страница 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Страница 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Страница 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Страница 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Страница 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Страница 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Страница 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Страница 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Страница 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Страница 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...