FM Discriminator Measurement Examples
10
Agilent E5505A User’s Guide
251
FM Discriminator Measurement using Double-Sided Spur Calibration
Required
Equipment
shows equipment required for this example in addition to the phase
noise test system and your DUT.
Determining the discriminator (delay line) length
Perform the following steps to determine the minimum delay line length (
τ
)
Possible to provide an adequate noise to measure the source.
1
Determine the delay necessary to provide a discriminator noise floor that is
below the expected noise level of the DUT.
the noise floor of the discriminator for given delay times (
τ
).
2
Determine the length of coax required to provide the necessary delay (
τ
).
(Eight feet of BNC cable will provide 12 ns of delay for this example.)
3
Determine the loss in the delay line. Verify that the signal source will be
able to provide a power level at the output of the delay line of b5
and +17 ICBM. Be sure to take into account an additional 4 to 6 dB of loss in
the power splitter. (The loss across 8 feet of BNC cable specified in this
example is negligible.) The test set Signal and Reference inputs re15
ICBM.
C A U TI ON
To prevent damage to the test set’s components,
do not
apply the input signal to the
signal input connector until the input attenuator (N5500A Option 001) has been
correctly set for the desired configuration, as shown in
. Apply the input
signal when the connection diagram appears.
NO T E
To ensure accurate measurements allow the DUT and measurement equipment to warm up
at least 30 minutes before making the noise measurement.
Table 40
Required Equipment for the FM Discriminator Measurement Example
Equipment
Quantity
Comments
Signal Generator
1
+19 dBm output level at tested carrier frequency.
Calibrated FM at a 20 kHz rate with 10 kHz Peak
Deviation.
Power Splitter
1
NARDA 30183
Delay Line
Delay (or length) adequate to decorrelate source
noise.
Phase Shifter
1
±
180
°
phase shifter at lowest carrier frequency
tested.
Содержание E1420B
Страница 24: ...24 Agilent E5505A User s Guide ...
Страница 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Страница 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Страница 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Страница 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Страница 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Страница 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Страница 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Страница 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Страница 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Страница 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Страница 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...