388
Agilent E5505A User’s Guide
17
System Specifications
Specifications
This section contains mechanical and environmental specifications, operating
characteristics, power requirements, and PC requirements for the system. It
also provides specifications for accuracy, measurement qualifications, and
tuning.
contains the mechanical and environmental specifications for a
system.
shows the system’s operating characteristics.
Table 66
Mechanical and environmental specifications
Specifications
Values
Temperature:
Operating
Nonoperating/storage
05
°
C to 40
°
C (41
°
F to 104
°
F)
-40
°
C to 70
°
C (-40
°
F to 158
°
F)
Altitude
Up to 2,000 m (6,500 ft)
Maximum relative humidity
80% for temperatures up to 31
°
C, decreasing linearly to 50% relative
humidity at 40
°
C
Air flow space required
102 mm (4 in) on all sides
System weight:
Benchtop
1.6 Meter Rack
2 Meter Rack
Approximate, typical
210 lbs (95.5 kg)
431 lbs (195.5 kg)
465 lbs (211 kg)
System dimensions:
Height
Width
with rack-mounted flat-panel display
with benchtop PC and display
Depth
711 to 915 mm (28 to 36 in)
432 mm (17 in)
n/a
1,016 mm (40 in)
711 mm (28 in)
1,620 mm (64 in)
600 mm (23.6 in)
1438 mm (56.7 in)
n/a
905 mm (35.7 in)
2,020 mm (79.5 in)
600 mm (23.6 in)
1438 mm (56.7 in)
n/a
905 mm (35.7 in)
Table 67
Operating characteristics
Warm up time required
30 minutes
Carrier frequency ranges
50 KHz to 1.6 GHz
50 KHz to 6 GHz
50 KHz to 18 GHz
50 KHz to 26.5 GHz
Offset frequency ranges
0.01 Hz to 2 MHz
0.01 Hz to 100 MHz
System noise response
–180 dBc/Hz typically (>10 kHz offsets)
System spurious response
≤
120 dBc typically
Содержание E1420B
Страница 24: ...24 Agilent E5505A User s Guide ...
Страница 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Страница 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Страница 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Страница 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Страница 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Страница 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Страница 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Страница 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Страница 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Страница 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Страница 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...