FM Discriminator Fundamentals
9
Agilent E5505A User’s Guide
245
The double-balanced mixer, acting as a phase detector, transforms the
instantaneous phase fluctuations into voltage fluctuations (
). With the
two input signals 90
°
out of phase (phase quadrature), the voltage out is
proportional to the input phase fluctuations. The voltage fluctuations can then
be measured by the baseband analyzer and converted to phase noise units.
The discriminator transfer response
The important equation is the final magnitude of the transfer response.
Where
represents the voltage fluctuations out of the discriminator and
represents the frequency fluctuations of the DUT.
is the phase
detector constant (phase to voltage translation).
is the amount of delay
provided by the delay line and
is the frequency offset from the carrier that
the phase noise measurement is made.
System sensitivity
A frequency discriminator’s system sensitivity is determined by the transfer
response. As shown below, it is desirable to make both the phase detector
constant
and the amount of delay
large so that the voltage fluctuations
out of a frequency discriminator will be measurable for even small
fluctuations .
The magnitude of the sinusoidal output term or the frequency discriminator is
proportional to
. This implies that the output response will have
peaks and nulls, with the first null occurring at
. Increasing the rate of
a modulation signal applied to the system will cause nulls to appear at
frequency multiples of
(
∆φ
∆
V
→
∆
V fm
( )
K
φ
2
πτ
d
∆
f fm
( )
π
fm
τ
d
(
)
sin
π
fm
τ
d
(
)
-----------------------------
=
∆
V fm
( )
∆
f fm
( )
K
φ
τ
d
fm
K
φ
τ
d
∆
V
∆
f
∆
V fm
( )
K
φ
2
πτ
d
π
fm
τ
d
(
)
sin
π
fm
τ
d
(
)
-----------------------------
∆
f fm
( )
(
)
=
NO T E
The system sensitivity is independent of carrier frequency .
fo
π
fm
τ
d
(
) π
fm
τ
(
⁄
n
m
1
τ
⁄
=
1
τ
d
⁄
Содержание E1420B
Страница 24: ...24 Agilent E5505A User s Guide ...
Страница 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Страница 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Страница 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Страница 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Страница 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Страница 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Страница 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Страница 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Страница 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Страница 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Страница 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...