270
Agilent E5505A User’s Guide
10
FM Discriminator Measurement Examples
7
Choose the
Sources
tab from the
Define Measurement
window.
a
Enter the carrier (center) frequency of your DUT (5 MHz to 1.6 GHz).
Enter the same frequency for the detector input frequency. See
.
.
8
Choose the
Cal
tab from the
Define Measurement
window.
9
Select
Derive constant from FM rate and deviation
as the calibration
method.
10
Set the
FM Rate
to 20 kHz and
FM Deviation
to 10 kHz, which are the
recommended FM rate and deviation. See
Figure 202
Enter frequencies in Source tab
e5505a_user_enter_source_tab
25 Jun 04 rev 2
Содержание E1420B
Страница 24: ...24 Agilent E5505A User s Guide ...
Страница 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Страница 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Страница 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Страница 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Страница 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Страница 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Страница 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Страница 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Страница 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Страница 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Страница 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...