Residual Measurement Fundamentals
7
Agilent E5505A User’s Guide
213
Measured ± DC peak voltage
Advantages
•
Easy method for calibrating the measurement system.
•
This calibration technique can be performed using the baseband analyzer.
•
Fastest method of calibration. If, for example, the same power levels are
always at the phase detector, as in the case of leveled, or limited outputs,
the phase detector sensitivity will always be essentially equivalent (within
one or two dB). Recalibration becomes unnecessary if this accuracy is
adequate.
•
Only one RF source is required.
•
Measures the phase detector gain in the actual measurement configuration.
This technique requires you to adjust off of quadrature to both the positive
and the negative peak output of the Phase Detector. This is done by either
adjusting the phase shifter or the frequency of the source. An oscilloscope
or voltmeter can optionally be used for setting the positive and negative
peaks.
Disadvantages
•
Has only moderate accuracy compared to the other calibration methods.
•
Does not take into account the amount of phase detector harmonic
distortion relative to the measured phase detector gain, hence the phase
detector must operate in its linear region.
•
Requires manual adjustments to the source and/or phase shifter to find the
phase detector’s positive and negative output peaks. The system will read
the value of the positive and negative peak and automatically calculate the
mean of the peak voltages which is the phase detector constant used by the
system.
Procedure
1
Connect circuit as per
on page 214, and tighten all connections.
2
Measure the power level that will be applied to the Signal Input port of the
test set’s phase detector.
on page 214 shows the acceptable
amplitude ranges for the E5505A system phase detectors.
Содержание E1420B
Страница 24: ...24 Agilent E5505A User s Guide ...
Страница 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Страница 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Страница 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Страница 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Страница 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Страница 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Страница 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Страница 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Страница 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Страница 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Страница 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...