Expanding Your Measurement Experience
4
Agilent E5505A User’s Guide
69
Selecting loop suppression verification
1
Using
as a guide, navigate to the
Cal
tab.
2
Check
Verify calculated phase locked loop suppression
and
Always Show
Suppression Graph
. Select
If limit is exceeded: Show Loop Suppression
Graph.
3
When you have completed these operations, click the
Close
button.
Setting up for the 8663A
10 MHz measurement
The signal amplitude at the test set’s R input (Signal Input) port sets the
measurement noise floor level. Use
page 70 to help determine the amplitude required to provide a noise floor level
that is below the expected noise floor of your DUT. For more information
about this graph, refer to
Figure 39
Selecting loop suppression verification
e5505a_user_select_loop
24 Jun 04 rev 3
Содержание E1420B
Страница 24: ...24 Agilent E5505A User s Guide ...
Страница 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Страница 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Страница 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Страница 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Страница 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Страница 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Страница 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Страница 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Страница 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Страница 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Страница 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...