246
Agilent E5505A User’s Guide
9
FM Discriminator Fundamentals
To avoid having to compensate for sin (x)/x response, measurements are
typically made at offset frequencies (
) much less
. It is possible to
measure at offset frequencies out to and beyond the null by scaling the
measured results using the transfer equation. However, the sensitivity of the
system get very poor results near the nulls.
The transfer function shows that increasing the delay
increases the
sensitivity of the system. However, increasing
also decreases the offset
frequencies (
) that can be measured without compensating for the sin(x)/x
response. For example, a 200 ns delay line will have better sensitivity close to
carrier than a 50 ns delay line., but will not be usable beyond 2.5 MHz offsets
without compensating for the sin(x)/x response; the 50 ns line is usable to
offsets of 10 MHz.
Increasing the delay
, also increases the attenuation of the line. While this
has no direct effect on the sensitivity provided by the delay line, it does reduce
the signal into the phase detector and can result in decreased
and
decreased system sensitivity.
The phase detector constant
equals the slope of the mixer sine wave output
at the zero crossings. When the mixer is not in compression,
equals
where
is the mixer efficiency and
is the voltage into the Signal Input
port (R port) of the mixer.
is also the voltage available at the output of the
delay line.
Figure 179
Nulls in sensitivity of delay line discriminator
e5505a_user_nulls_delay_line.ai
rev2 10/23/03
fm
1 2
τ
d
⁄
τ
d
τ
d
fm
τ
d
K
φ
K
φ
K
φ
K
L
V
R
K
L
V
R
V
R
Содержание E1420B
Страница 24: ...24 Agilent E5505A User s Guide ...
Страница 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Страница 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Страница 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Страница 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Страница 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Страница 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Страница 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Страница 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Страница 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Страница 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Страница 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...