AM Noise Measurement Fundamentals
11
Agilent E5505A User’s Guide
295
4
Reconnect the AM detector and enter the carrier-to-sideband ratio and
modulation frequency.
5
Measure the AM detector calibration constant (
.
6
Turn off AM.
7
Measure noise data and interpret the results.
NO T E
The carrier-to-sideband ratio
for AM is:
C
sb
-----
C
sb
-----
20
percentAM
100
-----------------------------
log
6
dB
=
=
Figure 228
Measuring the carrier-to-sideband ratio
Figure 229
Measuring the calibration constant
E5505a_meas_car_side_ratio
02 Mar 04 rev 1
Source
Modulation
analyzer
DUT with AM
E5505a_meas_cal_constant
02 Mar 04 rev 1
AM detector
Noise
input
Test set
NO T E
The quadrature meter should be at zero volts due to the blocking capacitor at the AM
detector’s output.
Содержание E1420B
Страница 24: ...24 Agilent E5505A User s Guide ...
Страница 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Страница 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Страница 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Страница 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Страница 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Страница 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Страница 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Страница 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Страница 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Страница 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Страница 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...