Expanding Your Measurement Experience
4
Agilent E5505A User’s Guide
85
Setting up the 8663A 10 MHz measurement
The signal amplitude at the R input (Signal Input) port on the test set sets the
measurement noise floor level. Use the graph in
and the example in
on page 86 to determine the amplitude. For more information, refer
to
, “Reference Graphs and Tables.”
Figure 51
Selecting loop suppression verification
e5505a_user_select_loop
24 Jun 04 rev 3
Figure 52
Noise floor for the 8644B 10 MHz measurement
-140
-160
-170
-180
-150
Expected phase noise floor of system (dBc/Hz)
f
10kHz
R Port signal level (dBm)
-15
+15
-5
+5
n5505a_exp_phase_noise
25 Feb 04 rev 1
L Port level +15dBm
Содержание E1420B
Страница 24: ...24 Agilent E5505A User s Guide ...
Страница 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Страница 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Страница 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Страница 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Страница 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Страница 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Страница 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Страница 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Страница 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Страница 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Страница 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...