Residual Measurement Examples
8
Agilent E5505A User’s Guide
239
3
Press the
Continue
button when ready to calibrate the measurement.
4
Adjust the phase difference at the phase detector as prompted by the phase
noise software. See
.
.
5
The system will measure the positive and negative peak voltage of the phase
detector using an internal voltmeter. The quadrature meter’s digital display
can be used to find the peak. The phase may be adjusted either by varying
the frequency of the source or by adjusting a variable phase shifter or line
stretcher.
6
The system software will then prompt you to set the phase noise software’s
meter to quadrature by adjusting the phase shifter until the meter indicates
0 volts, then press Continue. See
Figure 175
Adjust phase difference at phase detector
e5505a_user_adjust_dif_phase
24 Jun 04 rev 3
Содержание E1420B
Страница 24: ...24 Agilent E5505A User s Guide ...
Страница 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Страница 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Страница 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Страница 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Страница 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Страница 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Страница 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Страница 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Страница 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Страница 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Страница 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...