Residual Measurement Fundamentals
7
Agilent E5505A User’s Guide
209
User entry of phase detector constant
This calibration option requires that you know the phase detector constant for
the specific measurement to be made. The phase detector constant can be
estimated from the source power levels (or a monitor oscilloscope) or it can be
determined using one of the other calibration methods.
Once determined, the phase detector constant can be entered directly into the
system software without going through a calibration sequence. Remember,
however, that the phase detector constant is unique to a particular set of
sources, the RF level into the phase detector and the test configuration.
Advantages
•
Easy method for calibrating the measurement system.
•
Requires little additional equipment: only an RF power meter to manually
measure the drive levels into the phase detector or monitor oscilloscope.
•
Fastest method of calibration. If the same power levels are always at the
phase detector, (as in the case of leveled outputs), the phase detector
sensitivity will always be essentially the same (within a dB or two). If this
accuracy is adequate, it is not necessary to recalibrate.
•
Only one RF source is required.
•
Super-quick method of estimating the phase detector constant and noise
floor to verify other calibration methods and check available dynamic
range.
Disadvantages
•
The user entry of the phase detector constant is the least accurate of all the
calibration methods.
•
It does not take into account the amount of power at harmonics of the
signal.
Содержание E1420B
Страница 24: ...24 Agilent E5505A User s Guide ...
Страница 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Страница 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Страница 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Страница 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Страница 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Страница 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Страница 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Страница 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Страница 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Страница 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Страница 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...