AM Noise Measurement Examples
12
Agilent E5505A User’s Guide
309
shows the system interconnections.
Table 47
Test set signal input limits and characteristics
Limits
Frequency
•
50 kHz to 1.6 GHz (Std.)
•
50 kHz to 26.5 GHz (Option 001)
•
50 kHz to 26.5 GHz (Option 201)
Maximum Signal Input Power
Sum of the reference and signal input power
shall not 23 dBm
At Attenuator Output, Operating Level Range:
•
RF Phase Detectors
•
0 to +23 dBm (Signal Input)
•
+15 to +23 dBm (Reference Input)
•
Microwave Phase Detectors
•
0 to +5 dBm (Signal Input)
•
+7 to +10 dBm (Reference Input)
•
Internal AM Detector
0 to +20 dBm
Downconverters:
•
Agilent N5502A/70422A
+5 to +15 dBm
•
Agilent N5507A/70427A
0 to +30 dBm
Characteristics
Input Impedance
50
Ω
nominal
AM Noise
DC coupled to 50
Ω
load
Figure 247
Connect diagram example
+23 dBm
1.2-26.5 GHz
+10 dBm
50 kHz -1600 MHz
MAXIMUM POWER
20 mA MAX
TO
DOWNCONVERTER
FROM
DOWNCONVERTER
TUNE VOLTAGE
OUT OF LOCK
<100 kHz
ANALYZER
ANALYZER
RF ANALYZER
PHASE DET OUTPUT
MONITOR
+15 dBm MIN
50 kHz - 1600 MHz
+7 dBm MIN
1.2 - 26.5 GHz
REF INPUT
ERR
ACT
STATUS
SRQ
TLK
LSN
GPIB
RMT
NOISE
0.01 Hz - 100 MHz
1 V Pk
50
50 kHz - 26.5 GHz
INPUT
SIGNAL
POSSIBLE
SIGNAL INPUT
0 VDC MAX
<100 MHz
50
50 kHz -1600 MHz
+23 dBm + ATTEN
1.2 GHz-26.5 GHz
+10 dBm + ATTEN
+30 dBm MAX WITH
ATTENUATOR
MAXIMUM POWER
+30 dBm
OUTPUT POWER
POWER
N5500A Opt 001
Test Set
OUTPUT
RF ANALYZER
1.8-18 GHz
AUX LO
IF
5 -1500 MHz
ANALYZER
FROM TEST SET
10 VOLTS MAX
CONTROL
VOLTAGE
5 MHz -18 GHz
+5-
+15 dBm
SIGNAL
INPUT
RMT LSN TLK SRQ
ERR
ACT
STATUS
GPIB
5 MHz-1 GHz
+10 dBm
1 GHz-18 GHz
+15 dBm
MAXIMUM POWER
0 VDC MAX
SIGNAL INPUT
POWER
N5502A
Downconverter
Test set
Opt. 001
Downconverter
e5505a_user_sys_interconnect_dia
27 Jun 04 rev 1
Spectrum analyzer
Digitizer
input
Digitizer
output
GPIB
Display
PC
To test set rear panel
CHIRP input
DUT
Содержание E1420B
Страница 24: ...24 Agilent E5505A User s Guide ...
Страница 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Страница 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Страница 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Страница 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Страница 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Страница 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Страница 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Страница 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Страница 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Страница 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Страница 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...