![Agilent Technologies AssayMAP Bravo Скачать руководство пользователя страница 94](http://html1.mh-extra.com/html/agilent-technologies/assaymap-bravo/assaymap-bravo_installation-manual_2867689094.webp)
Verifying the installation and setup
Washing the syringes to prepare for first time use
86
AssayMAP Bravo Platform Installation Guide
5
To recover from the emergency stop:
a
At the emergency- stop pendant, turn the red button clockwise. The
spring- loaded button pops up.
b
In the
AssayMAP Bravo Error
dialog box, click
Abort
to end the protocol.
c
Run the Startup utility again to reset the system. For details, see
“Running the Startup utility” on page 83
.
6
Click
Run Cartridge Transfer
to start another run.
7
As soon as the Bravo head starts to move away from its position above the
wash station and towards deck location 6, use a pencil to interrupt the
Light Curtain.
The Bravo head stops moving and an error message appears.
8
To recover from the emergency stop:
a
Ensure that no objects are in the path of the Light Curtain.
b
In the
AssayMAP Bravo Error
dialog box, click
Abort
to end the protocol.
c
Run the Startup utility again to reset the system. For details, see
“Running the Startup utility” on page 83
.
Washing the syringes to prepare for first time use
Use the following procedure to prepare the new syringes for use. The
procedure primes and washes the syringes for 60 cycles.
Before you start
Ensure that you have the following required materials.
Item
Name (manufacturer and part number)
One
labware
for wash
buffer
Use one of the following labware:
•
Reservoir, Seahorse 201254- 100, PP, no walls, pyramid
bottom (Agilent 201254- 100)
•
12- Column Low- Profile Reservoir (Agilent 201280- 100)
•
8- Row Low- Profile PolyPro Reservoir (Agilent 201282- 100)
Deionized
water
Not applicable