![Agilent Technologies AssayMAP Bravo Скачать руководство пользователя страница 108](http://html1.mh-extra.com/html/agilent-technologies/assaymap-bravo/assaymap-bravo_installation-manual_2867689108.webp)
Manually updating teachpoints in each profile
Workflow for setting teachpoints in each profile manually
100
AssayMAP Bravo Platform Installation Guide
Workflow for setting teachpoints in each profile manually
About this topic
Every device profile in the Protein Sample Prep Workbench must be adjusted
to the correct teachpoints of the specific AssayMAP Bravo Platform. You can
use the following workflow to set the teachpoints for all nine deck locations in
every device profile manually.
Alternatively, for a semi- automated workflow, see
teachpoints in AssayMAP 1 profile” on page 58
Before you start
Ensure that all device profiles in the Protein Sample Prep Workbench have
been installed, and that they are configured correctly. See
settings and configuration” on page 39
.
Workflow for manually setting teachpoints
CAUTION
Using a profile that is associated with teachpoints from a different instrument
can result in a hardware collision and cause equipment damage. Ensure that you verify the
teachpoints associated with each profile for your specific instrument and Bravo 96AM Head.
Step
For this task...
See...
1
Initialize the AssayMAP 1 profile.
“Initializing the profile that you are
verifying” on page 51
2
In the AssayMAP 1 profile, set the teachpoint for
deck location 2 and derive the teachpoints for the
other eight deck locations automatically.
“Teaching part 1: Setting
teachpoints in AssayMAP 1 profile”
on page 58
3
Fine tune the derived teachpoints.
“Fine tuning the teachpoints” on
page 61
4
Normalization profile only
. If the Bravo Plate Riser
is installed at deck locations 2 and 6, adjust the
teachpoint at these locations to accommodate the
extra height.
“Adjusting the teachpoint for the
Bravo Plate Riser” on page 101
5
GlykoPrep- Plus profile only
. If the Peltier Thermal
Station is installed at deck location 4, adjust the
teachpoint to accommodate the red PCR plate insert.
“Adjusting the teachpoint for the
red PCR plate insert” on page 103
6
Orbital Shaking Station only
. If the accessory is
installed at deck location 9, fine tune the teachpoint.
“Teaching the Orbital Shaking
Station at deck location 9” on
page 62
7
Adjust the teachpoint for the 96AM Wash Station.
“Teaching the 96AM Wash Station
at deck location 1” on page 63