Verifying the installation and setup
Verification workflow
80
AssayMAP Bravo Platform Installation Guide
Verification workflow
After the AssayMAP Bravo Platform is set up in the VWorks software, perform
the following steps in the order given to verify the installation and setup.
About installing the Bravo pendant, Light Curtain, and shields
Ensure that the following are installed properly. For instructions, see the
Bravo Platform Safety and Installation Guide
.
•
Emergency- stop pendant.
Ensure that the pendant and its cable are in
good condition and connected to the Pendant port on the junction box.
•
Light Curtain
. Ensure the lightpost transmitter (TX) and receiver (RX)
cables are connected and that the light beams are aligned correctly.
•
Bravo front, side and rear shields
. Ensure that the accessory tubing and
cabling is routed off the Bravo deck through the access windows in the
side and rear shields.
Step
For this task...
See...
1
Ensure that the Light Curtain,
pendant, and shields are
installed.
“About installing the Bravo
pendant, Light Curtain, and
shields” on page 80
2
Open the Protein Sample Prep
Workbench and access the
Utility Library.
“Opening the Workbench and
accessing the utilities” on page 82
3
Run the Startup utility.
“Running the Startup utility” on
page 83
4
Test the safety equipment.
“Testing the safety equipment” on
page 84
5
Run the Syringe Wash utility
for 60 cycles.
“Washing the syringes to prepare
for first time use” on page 86
6
Perform a mock run of the
Affinity Purification protocol.
“Performing a mock run” on
page 90
7
Run the Syringe Test utility.
“Running the Syringe Test utility”
on page 97
8
Run the Shutdown utility.
“Running the Shutdown utility” on
page 98
WARNING
Moving-parts hazards are accessible through the front, rear, and side
openings of the Bravo Platform. To prevent potential injury, ensure that the safety
equipment is installed and in good working condition.