Manually updating teachpoints in each profile
Adjusting the teachpoint for the red PCR plate insert
AssayMAP Bravo Platform Installation Guide
8
Repeat this procedure for the other deck location where the Bravo Plate
Riser will be installed.
Verifying the teachpoint for the Bravo Plate Riser
Use the following procedure to verify the adjusted teachpoint for the Bravo
Plate Riser.
1
Ensure that the Bravo Plate Riser is securely seated on the platepad at the
corresponding deck location. Ensure that the Bravo Plate Riser is oriented
so that the crosshairs is at the back left corner.
2
In the
Jog/Teach
tab, click the
Location
list, and then select the deck
location you are teaching.
3
Click
Move
, The Bravo head moves to the selected deck location so that the
tip is at the crosshairs on the Bravo Plate Riser.
4
Visually inspect the alignment to ensure that the tip is centered from left
to right and front to back.
Adjusting the teachpoint for the red PCR plate insert
About this topic
Use the following procedure only if the deck configuration includes the Peltier
Thermal Station and you plan to use the red PCR plate insert.
The AssayMAP GlykoPrep- plus applications require the red PCR plate insert to
be present in the plate nest of the Peltier Thermal Station for the entire
workflow. Use the following teaching procedure to accommodate this
requirement for deck location 4 in the GlykoPrep- plus profile.
Figure
PCR plate insert
IMPORTANT
Ensure that the PCR plate insert is not installed in the Peltier
Thermal Station at deck location 4 for the following teaching procedure.
CAUTION
A hardware collision can occur if you initialize a device profile that does not
include the teachpoints for the Bravo Plate Riser while the Bravo Plate Riser remains on the
deck. Ensure that you remove the Bravo Plate Riser from the deck when initializing a different