![Agilent Technologies AssayMAP Bravo Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/agilent-technologies/assaymap-bravo/assaymap-bravo_installation-manual_2867689021.webp)
13
Installing the platform
Installing the 96AM Wash Station and Pump Module
AssayMAP Bravo Platform Installation Guide
Figure
Deck location 1 (top view) showing crosshairs orientation
2
On the platepad side that is next to the crosshairs, using a hex driver
uninstall the two 2- mm screws and remove the bracket.
Note:
The wash station is easier to install if you remove the outer
bracket, and then replace the bracket after the wash station is in
position.
Figure
Bracket next to platepad crosshairs
3
Install the wash station on the platepad as follows:
a
Ensuring that the inlet and outlet ports are towards the rear of the
deck, place the wash station on the platepad.
b
Guide the platepad’s inner bracket into the wash station’s horizontal
side slot.
c
Re- install the bracket that you removed in step 2 on the outer side of
the platepad. Ensure that you guide the bracket into the corresponding
horizontal slot on the wash station. Ensure that both brackets remain
engaged as you tighten the bracket screws.
d
Ensure that the wash station sits level on the platepad and that both
brackets are holding the wash station securely in place.
00240
Bravo
Plate Orientation
1
2
3
4
5
6
7
8
9
00240