69
Setting the teachpoints
Teaching part 3: Updating the other profiles based on AssayMAP 1
AssayMAP Bravo Platform Installation Guide
2
In the
Open
dialog box, ensure the file type is
Protocol Files (*.pro)
. Locate the
C:\VWorks Workspace\Workbench\Protocols folder, select the
AssayMAP
Teachpoint Update v1p0.pro
file, and then click
Open
.
3
When the following VWorks initialization message appears, click
Yes
.
The AssayMAP Bravo.dev file opens, and the AssayMAP 1 profile initializes.
4
Ensure that the AssayMAP Teachpoint Update protocol is open in the
VWorks window, and then click
Start
.