![Agilent Technologies AssayMAP Bravo Скачать руководство пользователя страница 112](http://html1.mh-extra.com/html/agilent-technologies/assaymap-bravo/assaymap-bravo_installation-manual_2867689112.webp)
Manually updating teachpoints in each profile
Adjusting the teachpoint for the red PCR plate insert
104
AssayMAP Bravo Platform Installation Guide
Adjusting the teachpoint for the PCR plate insert
To adjust the teachpoint for the PCR plate insert:
1
In the
Jog/Teach
tab, click
Location 4
.
2
Click
Move
to move the head to the previously defined teachpoint for
location 4.
3
Using the Jog Axes controls, jog the head up (- Z) by exactly 4.4 mm.
For example, if the value for the teachpoint’s Z- axis coordinate from the
fine tuning was 104.00 mm, the new value after adjusting by 4.4 mm will
be 99.60 mm.
4
Click
Teach
to set the new teachpoint.
5
To save the new teachpoint for this location, click the
Profiles
tab, and then
click
Update this profile.