Installing the platform
Installing the Peltier Thermal Station
22
AssayMAP Bravo Platform Installation Guide
Use the following workflow.
Installing the Peltier Thermal Station in the Bravo deck
Use the following procedure to install the Peltier Thermal Station at deck
location 4.
To install the Peltier Thermal Station:
1
Uninstall the platepad, if applicable, from deck location 4.
To uninstall a platepad, remove the M6 flathead screw from the center of
the platepad. Remove the platepad from the deck.
2
Remove the cutout cover. Use a hex driver to uninstall the four M3 screws
that attach the cutout cover to the deck.
Figure
Cutout cover on the Bravo deck
Step
For this task...
See...
1
Install the Peltier Thermal
Station.
“Installing the Peltier Thermal Station
in the Bravo deck” on page 22
2
Connect the STC Controller.