59
Setting the teachpoints
Teaching part 1: Setting teachpoints in AssayMAP 1 profile
AssayMAP Bravo Platform Installation Guide
IMPORTANT
Ensure that the pipette tip is new and in good condition.
•
0.010- inch (0.254 mm) feeler gauge
This gauge is used to set the vertical distance between the pipette tip and
top of the crosshairs mark on the platepad or accessory.
The following procedures assume that you have basic knowledge of the Bravo
teaching process and have been trained to teach the Bravo Platform.
Ensure the following:
•
All Workbench profiles have been configured correctly. See
profile settings and configuration” on page 39
•
The Protein Sample Prep Workbench software is not open.
•
In the VWorks software, Simulation is off.
T
Workflow for setting the teachpoints
Setting the teachpoint for deck location 2 and deriving the remaining teachpoints
To set the teachpoints:
1
Manually seat a 250- µL pipette tip on probe A1 (back left corner of the
Bravo 96AM Head). Ensure that the pipette tip locks into place with the
probe.
Ensure a standard platepad is installed at deck location 2.
CAUTION
To prevent potential equipment damage, ensure that the deck is clear of any
obstacles before initializing the AssayMAP Bravo Platform.
Step
For this task...
See...
1
In the
Bravo Diagnostics
dialog
box, initialize the AssayMAP 1
profile.
“Initializing the profile that you are
verifying” on page 51
2
Set the teachpoint for deck
location 2 and derive the
remaining teachpoints for the
AssayMAP 1 profile.
“Setting the teachpoint for deck
location 2 and deriving the
remaining teachpoints” on page 59
3
Adjust the accessory location
teachpoints for the
AssayMAP 1 profile.
“Teaching part 2: Adjusting
teachpoints for accessories” on
page 62
4
Use the AssayMAP Update
Teachpoints protocol to update
the remaining profiles.
“Teaching part 3: Updating the
other profiles based on
AssayMAP 1” on page 67
Содержание AssayMAP Bravo
Страница 1: ...AssayMAP Bravo Platform Installation Guide Original Instructions...
Страница 86: ...Setting the teachpoints Checking and adjusting the gripper offset 78 AssayMAP Bravo Platform Installation Guide...
Страница 113: ......
Страница 114: ...Agilent Technologies Installation Guide G5571 90001 Revision B March 2018...