![Agilent Technologies AssayMAP Bravo Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/agilent-technologies/assaymap-bravo/assaymap-bravo_installation-manual_2867689059.webp)
51
Setting up the computer and software
Initializing the profile that you are verifying
AssayMAP Bravo Platform Installation Guide
Initializing the profile that you are verifying
Initializing the profile establishes communication with the AssayMAP Bravo
Platform and accessories. You will initialize a profile to test the accessory
settings and verify the gripper adjustment for that profile.
Before you start
Initializing the profile in Bravo Diagnostics
To initialize the profile:
1
In the
Profiles
tab, ensure that the correct profile is selected, and then click
Initialize this profile
to start initializing the selected profile.
2
If the
There appears to be a plate present
error message appears, click
Ignore and
Continue
to continue the homing process.
3
When the
Verify that it is safe to home the W-axis
message appears, click
Retry
to
continue homing the
w
- axis.
WARNING
When you initialize the Bravo Platform, the head and tie bar can move. To
prevent potential injury, keep clear of the device while it is in motion.