Setting up the computer and software
Verifying the profile settings and configuration
40
AssayMAP Bravo Platform Installation Guide
5
In the
<name>
.dev tab, ensure that the
Profile
selection is correct.
If necessary, select the correct name from the
Profile
list, and then click the
Save
icon.
Figure
AssayMAP Bravo.dev tab and Profile setting for AssayMAP 1
6
In the
<name>
.dev tab, click
Device diagnostics
. The
Bravo Diagnostics
dialog
opens.
7
In the
Profiles
tab, verify the
Profile name
.
8
Verify the following
Head Information
properties.
Device file name
Profile name
AssayMAP Bravo.dev
AssayMAP 1
GlykoPrep- Plus_AM Bravo.dev
GlykoPrep- Plus
AM Peptide Sample Prep.dev
AssayMAP Peptide Sample Prep
InSolution Digestion- AM Bravo.dev
Single Plate InSolution Digestion
Normalization- AM Bravo.dev
AssayMAP Normalization