53
Setting up the computer and software
Testing the accessories
AssayMAP Bravo Platform Installation Guide
Testing the Pump Module
Use the following procedure to test the communication with the Pump Module.
In a subsequent verification procedure, you will run the Startup utility, which
will test the functioning of the Pump Module and wash station using liquid.
To test the communication with the Pump Module:
1
Ensure that the Pump Module is on.
2
Lift the flip- top cover on each pump head so that it is fully open.
Note:
While the pump heads are open, no liquid will pump in the
subsequent steps.
3
In the
Configuration
tab, click the
Autofill Station
icon, and then click
Diagnose
accessory
. The Bravo Accessory Diagnostics dialog box opens.
00306