Installing the platform
Installing the Bravo risers
6
AssayMAP Bravo Platform Installation Guide
2
Ensure that the head mount is resting against the Bravo back plate. With a
person positioned on each side of the Bravo Platform, slowly tilt the device
backwards until the back plate rests on the foam.
3
Remove the four rubber feet from the base of the Bravo deck.
Figure
Bottom view of Bravo deck
4
On the bottom of the Bravo deck, install the two risers, one on each side.
Use the four screws and washers provided to install each riser. Tighten the
screws as follows:
a
Tighten the screws at the top first, and then tighten the screws at the
bottom, as the following figures show.
b
Tighten the screws until snug, and ensure that the risers are flush to
the base of the Bravo deck.
Figure
Top screw located at inner hole
Содержание AssayMAP Bravo
Страница 1: ...AssayMAP Bravo Platform Installation Guide Original Instructions...
Страница 86: ...Setting the teachpoints Checking and adjusting the gripper offset 78 AssayMAP Bravo Platform Installation Guide...