4
Installing 7890 GC Columns
The GC/MSD Interface for a 7890B Series GC
5977B Series MSD Operating Manual
99
The GC/MSD interface is heated by an electric cartridge heater. Normally, the
heater is powered and controlled by Thermal Aux #2 heated zone of the GC. The
interface temperature can be set from Agilent MassHunter GC/MS Acquisition
software or from the GC. A thermocouple sensor in the interface monitors the
temperature.
The GC/MSD interface should be operated in the range of 250
to 350
C. The
interface temperature should be slightly higher than the maximum GC oven
temperature, but
never
higher than the maximum column temperature.
The GC/MSD interface can be used with both EI and CI sources. The EI XTR, HES,
and CI sources require a tip seal assembly. The tip seal, spring, and knurled tip
seal retainer interfere with, and must be removed when using the EI SS and Inert
sources.
“To Install a Capillary Column in the GC/MSD Interface Using the
“To Install a Capillary Column in
the GC/MSD Interface Using a Standard Column Nut”
WARNING
The GC/MSD interface operates at high temperatures. If you touch it when it is
hot, it will burn you.
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...