Installing 8890 GC Columns
To Install a Capillary Column in the GC/MSD Interface Using the Self-Tightening Column Nut
42
5977B Series MSD Operating Manual
To Install a Capillary Column in the GC/MSD
Interface Using the Self-Tightening Column
Nut
This procedure is for the installation of a capillary column directly into the
analyzer using the Agilent recommended self-tightening column nut.
Materials needed
•
Interface tip seal (G3870-20542) (See
on page 44.) (Not used with EI
SS or Inert source)
•
Knurled tip seal retainer (G3870-20547) (Not used with EI SS or Inert source)
•
Tip seal spring (G7005-20024) (Not used with EI SS or Inert source)
•
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
•
Flashlight
•
Magnifying loupe
•
Gloves, clean
•
Large (8650-0030)
•
Small (8650-0029)
•
Self Tightening column nut for GC\MSD interface (5190-5233)
•
Ferrules, Vespel
•
0.27 mm id, for 0.10 mm id columns (5062-3518)
•
0.37 mm id, for 0.20 mm id columns (5062-3516)
•
0.40 mm id, for 0.25 mm id columns (5181-3323)
•
0.5 mm id, for 0.32 mm id columns (5062-3514)
•
0.8 mm id, for 0.53 mm id columns (5062-3512)
•
Septum (may be old, used inlet septum)
•
Safety glasses
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...