Installing 8890 GC Columns
To Install a Capillary Column in the GC/MSD Interface Using the Self-Tightening Column Nut
46
5977B Series MSD Operating Manual
15
Gently
check the alignment of the ion source and the interface tip seal.
When the ion source is aligned correctly, the analyzer chamber can be
closed all the way with no resistance except the spring tension from the
interface tip seal.
16
You can align the ion source and interface tip seal by wiggling the side plate
on its hinge. If the door still will not close, contact your Agilent Technologies
service representative.
17
Close the analyzer chamber door. (See