3
Installing Intuvo 9000 GC Columns
To Replace an Intuvo GC Column
5977B Series MSD Operating Manual
59
To Replace an Intuvo GC Column
This procedure applies to GCs with a single column. For a 2 column replacement
see the
Agilent Intuvo 9000 Gas Chromatograph Maintaining Your GC
manual.
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Tweezers (8710-2460)
•
Intuvo gasket, polyimide 5/pk, for temperatures < 350 °C (5190-9072)
•
Intuvo gasket, nickel 5/pk, for temperatures 350 °C to 450 °C (5190-9073)
•
Intuvo torque driver stored on oven door (5190-9571)
Procedure
1
Prepare the GC for maintenance.
From the GC panel select
Maintenance > Column > Perform Maintenance >
Install Column > Start Maintenance
. This procedure cools the inlet, detector,
column, Guard Chip, and other components in the flow path heated zones to
< 40 °C and configures the GC. Follow the on-screen GC prompts.
2
From MassHunter Data Acquisition, Vent the MSD (See
3
If you are using hydrogen or other flammable gas as a carrier gas, close the
manual gas supply valve to the instrument before turning off the MSD power.
WARNING
The inlet, detector, bus components, and column can be hot enough to cause
burns. Cool heated zones to a safe handling temperature before continuing.
CAUTION
Always wear clean gloves while handling any parts that are in the flow path
going to the analyzer or parts that are located inside the analyzer.
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...