Operating in CI Mode
To Use Other Reagent Gases
5977B Series MSD Operating Manual
157
Isobutane CI
Isobutane (C
4
H
10
) is commonly used for CI when less fragmentation is desired
in the CI spectrum. This is because the proton affinity of isobutane is higher than
that of methane; hence, less energy is transferred in the ionization reaction.
Addition and proton transfer are the ionization mechanisms most often associated
with isobutane. The sample itself influences which mechanism dominates.
Ammonia CI
Ammonia (NH
3
) is commonly used for CI when less fragmentation is desired in
the CI spectrum. This is because the proton affinity of ammonia is higher than
that of methane; hence, less energy is transferred in the ionization reaction.
Because many compounds of interest have insufficient proton affinities,
ammonia CI spectra often result from the addition of NH
4
+ and then, in some
cases, from the subsequent loss of water. Ammonia reagent ion spectra have
principal ions at
m/z
18, 35, and 52, corresponding to NH
4
+, NH
4
(NH
3
)+, and
NH
4
(NH
3
)
2
+.
To adjust your MSD for isobutane or ammonia CI, use the following procedure:
Procedure
1
Perform a standard PCI autotune with methane and PFDTD. (See
Perform a PCI Autotune (Methane Only)”
2
In the
Tune and Vacuum Control
view from the
Tune
menu, click
Tune
Wizard,
and when prompted, select
Isobutane
or
Ammonia
. This will change
the menus to use the selected gas and select appropriate default tune
parameters.
3
When prompted, select
Gas B
. (the port where Isobutane or Ammonia is
plumbed). Continue to the prompts from the Tune Wizard and set the gas
flow to 20%.
If you use an existing tune file, be sure to save it with a new name if you do
not want to overwrite the existing values. Accept the default temperature
and other settings.
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...