5
Operating in EI Mode
To Set Analyzer Temperatures from the Instrument Control View
5977B Series MSD Operating Manual
123
3
To send the new temperature parameters to the currently loaded tune file
and download these parameters to the MSD, click
Apply.
4
Click
Close
to exit the dialog. If changes were made to any parameters the
Save MS Tune File
dialog displays. Click
OK
to save your changes to the
same file, or type a new file name and click
OK
. Click
Cancel
to discard the
edit made to any parameter.