7
General Maintenance
Maintaining the Analyzer
5977B Series MSD Operating Manual
175
Some parts can be damaged by electrostatic discharge
The wires, contacts, and cables connected to the analyzer components can carry
electrostatic discharges (ESD) to the electronics boards to which they are
connected. This is especially true of the mass filter (quadrupole) contact wires
which can conduct ESD to sensitive components on the side board. ESD damage
may not cause immediate failure but will gradually degrade performance and
stability. (See
on page 171 for more information.)
Some analyzer parts should not be disturbed
The mass filter (quadrupole) requires no periodic maintenance. In general, the
mass filter should never be disturbed. In the event of extreme contamination, it
can be cleaned, but such cleaning should only be done by a trained Agilent
Technologies service representative. The HED ceramic insulator must never be
touched.
More information is available
If you need more information about the locations or functions of analyzer
components, refer to the
Agilent 5977B Series MSD Troubleshooting and
Maintenance Manual
.
Many procedures in this chapter are illustrated with video clips.
CAUTION
Electrostatic discharges to analyzer components are conducted to the side
board where they can damage sensitive components. Wear a grounded
antistatic wrist strap (see
other antistatic precautions
before
you open the analyzer chamber.
CAUTION
Incorrect handling or cleaning of the mass filter can damage it and have a
serious, negative effect on instrument performance. Do not touch the HED
ceramic insulator.
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...