7
General Maintenance
To Connect/Disconnect Wiring from the EI XTR, SS, and Inert Sources
5977B Series MSD Operating Manual
197
To Connect/Disconnect Wiring from the EI
XTR, SS, and Inert Sources
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Pliers, long-nose (8710-1094)
•
Tweezers (8710-2460)
Procedure
1
Use tweezers or needle nose pliers to connect/disconnect the ceramic
source board wire leads at the source connectors. (See
.)
2
Use tweezers or needle nose pliers to connect/disconnect the source heater
wire leads at the ceramic source board.
Figure 63 Wiring for the EI XTR Source
Extractor lens
Source heater
Repeller
Filament 1
Source RTD
Filament 2
Ceramic source board
Entrance lens
Ion focus lens