5
Operating in EI Mode
To Pumpdown the MSD
5977B Series MSD Operating Manual
141
10
In the
Instrument Control
view, from the Instrument menu, select
MS
Vacuum Control
to display the
Vacuum Control
11
Click
Pump Down
in the
Vacuum Control
dialog, and follow the system
prompts.
12
When prompted, turn on the GC/MSD interface heater and GC oven. Click
OK
when you have done so.
The software will turn on the ion source and mass filter (quad) heaters. The
temperature setpoints are stored in the current autotune (*.u) file.
13
After the message
Okay to run
appears, wait 2 hours for the MSD to reach
thermal equilibrium. Data acquired before the MSD has reached thermal
equilibrium may not be reproducible.
14
If using the JetClean system, prepare the hydrogen system for use. Refer to
the JetClean system Operating manual for warnings regarding hydrogen
safety.
Figure 48 Vacuum Control dialog
CAUTION
Do not turn on any GC heated zones until carrier gas flow is on. Heating a
column with no carrier gas flow will damage the column.
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...