7
General Maintenance
To Disassemble the EI HES
182
5977B Series MSD Operating Manual
9
If necessary, use gravity to remove the lens stack ceramic insulator from the
source body.
10
Remove the five lenses from the lens insulator/holder.
CAUTION
Use care when removing the lenses from the lens insulator casing. Putting
excessive stress on this casing can break or crack it. If this happens, do not
attempt to operate with a defective lens insulator, it must be replaced.
Figure 57 Exploded parts view of the EI HES
1
2
8
16
17
15
9
4
3
11
10
6
5
7
13
12
14
8
12
Table 18 Parts list for EI HES (
Item number
Item description
Part number
1
Source finger grip
G7002-20008
2
Filament block
G7002-20019
3
Extractor lens (5)*, with 3 mm opening
G7004-20061
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...