8
CI Maintenance
To Remove the EI HES Radiator
224
5977B Series MSD Operating Manual
To Remove the EI HES Radiator
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Tweezers (8710-2460)
•
T10 Torx screwdriver (5182-3466)
Procedure
1
Remove the EI HES. (See
2
Disconnect the wires from the EI HES radiator. (See
Wiring to the EI HES Radiator”
3
Use a T10 Torx screwdriver to loosen the 2 screws that secure the radiator to
the analyzer, and place the radiator in its storage container.
WARNING
The analyzers, GC/MSD interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.
CAUTION
•
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
•
Make sure you use an antistatic wrist strap and take other antistatic
precautions before touching analyzer components.
•
When disconnecting leads, pull on the connectors, not on the wires
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...