Operating in EI Mode
Controlling Column Flow
5977B Series MSD Operating Manual
115
Controlling Column Flow
Carrier gas flow is controlled by inlet pressure in the GC. Set the column mode to
Constant Pressure
to maintain a constant inlet pressure, and the column flow
will decrease as the GC oven temperature increases. With EPC and the column
mode set to
Constant Flow
, the same column flow is maintained regardless of
temperature.
The MSD can be used to measure actual column flow. You inject a
small
amount
of air or other unretained chemical, and time how long it takes to reach the MSD.
With this time measurement, you can calculate the column flow. (See
Calibrate Column Flow Linear Velocity in a 7890 GC”
on page 95 as appropriate.
For an 8890 GC, see the 8890 GC Operating Manual.)
Controlling JetClean Hydrogen Flow
The JetClean option uses hydrogen gas for cleaning the ion source. The
JetClean system MFC sends hydrogen to the to the analyzer through the CI line
in the GC/MSD interface. The flow rate is controlled by the MassHunter GC/MS
Acquisition software. The JetClean system is only supported on turbo based
systems using an EI Inert source, or an EI HES. (See
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...