5
Operating in EI Mode
To Set the GC/MSD Interface Temperature from MassHunter
124
5977B Series MSD Operating Manual
To Set the GC/MSD Interface Temperature
from MassHunter
Procedure
1
From
Instrument Control
view select
Instrument>GC Edit Parameters.
2
Click the
Aux Heaters
icon to edit the interface temperature. (See
.)
3
Select
On
to turn on the heater and type the setpoint in the
Value °C
column.
The typical setpoint is 280 °C.
4
Click
Apply
to download setpoints, or click
OK
to download setpoints and
close the window.
5
To make the new settings part of the method, select
Save
from the Method menu.
Figure 38 8890 GC edit parameters window
CAUTION
Ensure that the carrier gas is turned on and the column has been purged of air
before heating the GC/MSD interface or the GC oven.
When setting the GC/MSD interface temperature, never exceed the maximum
for your column.
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...